Measurements Failing After a Software Upgrade

Measurements Failing After a Software Upgrade

Overview

If measurements stop working after updating iCON Trades to a new version, follow the troubleshooting steps below in order. In some cases, this symptom is caused by a coincidental hardware issue (auto focus failure) rather than the software update itself.

Troubleshooting Steps

Step 1: Sensor Reset

  1. Press and hold the power button on the iCS instrument for 8 seconds.
  2. Allow the instrument to restart fully.
  3. Attempt a measurement again.

Step 2: Recreate the Device Profile

  1. Open Device Manager in iCON Trades.
  2. Delete the current device profile for the iCS.
  3. Create a new device profile for the instrument.
  4. Attempt a measurement again.

Step 3: Factory Reset the Instrument

  1. In iCON Trades, open the device profile for your iCS.
  2. Navigate to Info.
  3. Select Factory Reset and confirm.
  4. After the reset completes, recreate the device profile.

Step 4: Reset Calibration

  1. Open the device profile > Info.
  2. Select Calibration Reset and confirm.
  3. Test measurements again.

Step 5: Collect Logs and Contact Support

If none of the above steps resolve the issue, generate diagnostic logs while the iCS is connected to iCON Trades and send them to Advanced Dimensions support. In some cases, log analysis has revealed an underlying auto focus hardware failure that happened to coincide with the software update — this requires an RGA (Return Goods Authorization).

Related Articles

    • Related Articles

    • Installing and Reinstalling the Leica 3D Disto Software

      Overview The Leica 3D Disto software runs on Windows and must be installed on any PC or tablet you intend to use for measurements. This guide covers a fresh installation, reinstallation after a hardware change, and what to do if the software fails to ...
    • Upgrading from Leica 3D Disto Software v5 to v6

      Overview Version 6 of the Leica 3D Disto software introduced significant new features including Layers and CAD Tools arcs. However, upgrading from v5 to v6 is not a free update — it requires purchasing a new Customer Care Package (CCP) that includes ...
    • 3D Disto Software: What's New in Version 6 vs. Version 5?

      Overview Leica released Version 6.0 of the 3D Disto software for Windows as a paid upgrade from Version 5. If your customers are asking about upgrading, the two headline features are layer functionality and CAD tools arcs. This article explains what ...
    • Auto Focus Not Working / Unable to Capture vTargets or Use vPen

      Overview If your iCON iCS cannot capture vTargets or the vPen is not functioning despite good lighting conditions, the instrument may have an auto focus hardware failure. Follow the diagnostic procedure below before contacting support. Auto Focus ...
    • Activating a Software License on the iCON iCS

      Overview The iCON iCS requires a software license to be activated through the Device Manager on the tablet. An active internet connection is required to complete activation. Steps to Activate Open the iCON Trades app on your tablet. Navigate to Menu ...